| Nominal and guaranteed values | XS3DU Dual Range |
| Maximum capacity | 3.1 g |
| Readability | 0.01 mg |
| Maximum capacity fine range | 0.8 g |
| Readability fine range | 0.001 mg |
| Repeatability std. range - at nominal load | 0.006 mg |
| - at low load (measured at) | 0.005 mg (0.2 g) |
| Repeatability fine range - at nominal load | 0.001 mg |
| - at low load (measured at) | 0.0008 mg (0.2 g) |
| Linearity | 0.004 mg |
| Eccentric load deviation (test load)1) | 0.005 mg (2 g) |
| Techical data - typical values4) |
|
| Repeatability (sd) | 0.0005 mg +1.2 x (10–7)•R_gr |
| Differential linearity deviation (sd) | √2x(10-12)g·R_nt |
| Differential eccentric load deviation (sd) | 1.2 x (10–6)•R_nt |
| Sensitivity offset (sd)2) | 3 x (10–6)•R_nt |
| Minimum weight* (@ U=1 %, 2 sd) | 0.1 mg+2.4 x (10–6)•R_gr |
| Settling time standard range | < 6 sec |
| Settling time fine range | < 10 sec |
1)According to OIML R76
2)In the temperature range 10…30°C
3)Stability of sensitivity as from first installation with FACT
4)Can be used for the estimation of uncertainty sd: standard deviation Rgr: gross weight Rnt: net weight (sample weight) a: year (annum)
*Repeatability and minimum weight can be improved and affected by the following measures: - choice of suitable weighing parameters, - moving to better location, - using smaller tare containers