| Nominal and guaranteed values | XP6 |
| Maximum capacity | 6.1 g |
| Readability | 0.001 mg |
| Repeatability - at nominal load | 0.0008 mg |
| - at low load (measured at) | 0.0006 mg (0.2 g) |
| Linearity | 0.004 mg |
| Eccentric load deviation (test load)1) | 0.005 mg (2 g) |
| Sensitivity offset | 7 x 10-6 |
| Sensitivity temperature drift2) | 0.0001 %/°C |
| Sensitivity stability3) | 0.0001 %/a |
| Techical data - typical values4) |
| Repeatability (sd) | 0.0004 mg+3 x (10–8)•R_gr |
| Differential linearity deviation (sd) | √1.5x(10-13)g·R_nt |
| Differential eccentric load deviation (sd) | 5 x (10–7)•R_nt |
| Sensitivity offset (sd)2) | 1.5x(10-6)·R_nt |
| Minimum weight* (according to USP) | 1.2 mg+9 x (10–5)•R_gr |
| Minimum weight* (@ U=1 %, 2 sd) | 0.08 mg+6 x (10–6)•R_gr |
| Settling time | < 8 s |
1)According to OIML R76
2)In the temperature range 10…30°C
3)Stability of sensitivity as from first installation with FACT
4)Can be used for the estimation of uncertainty sd: standard deviation Rgr: gross weight Rnt: net weight (sample weight) a: year (annum)
*Repeatability and minimum weight can be improved and affected by the following measures: - choice of suitable weighing parameters, - moving to better location, - using smaller tare containers